Published at : 24 Dec 2024
Volume : IJtech
Vol 15, No 6 (2024)
DOI : https://doi.org/10.14716/ijtech.v15i6.7197
Adrian Nacarino | 1. Instituto de Investigación en Ciencias Biomédicas (INICIB), Ricardo Palma University, Av. Alfredo Benavides 5440, Santiago de Surco, 15039, Peru, 2. Professional School of Mechatronics Engineering, |
Anderson La-Rosa | Professional School of Mechatronics Engineering, Ricardo Palma University, Av. Alfredo Benavides 5440, Santiago de Surco, 15039, Peru |
Yelmo Quispe | Professional School of Mechatronics Engineering, Ricardo Palma University, Av. Alfredo Benavides 5440 Santiago de Surco, 15039, Peru |
Karl Castro | Professional School of Mechatronics Engineering, Ricardo Palma University, Av. Alfredo Benavides 5440 Santiago de Surco, 15039, Peru |
Freedy Sotelo Valer | Professional School of Mechatronics Engineering, Ricardo Palma University, Av. Alfredo Benavides 5440 Santiago de Surco, 15039, Peru |
Jose Cornejo | 1. Instituto de Investigación en Ciencias Biomédicas (INICIB), Ricardo Palma University, Av. Alfredo Benavides 5440, Santiago de Surco, 15039, Peru. 2. Research Group of Advanced Robotics and Mechatro |
Mariela Vargas | Instituto de Investigación en Ciencias Biomédicas (INICIB), Ricardo Palma University, Av. Alfredo Benavides 5440 Santiago de Surco, 15039, Peru |
Robert Castro | Professional School of Mechatronics Engineering, Ricardo Palma University, Av. Alfredo Benavides 5440 Santiago de Surco, 15039, Peru |
Ricardo Palomares | 1. Professional School of Mechatronics Engineering, Ricardo Palma University, Av. Alfredo Benavides 5440 Santiago de Surco, 15039, Peru. 2. Research Group of Advanced Robotics and Mechatronics (GI-RO |
Bryan Sanchez | Professional School of Mechatronics Engineering, Ricardo Palma University, Av. Alfredo Benavides 5440, Santiago de Surco, 15039, Peru |
David Allcca | Professional School of Mechatronics Engineering, Ricardo Palma University, Av. Alfredo Benavides 5440, Santiago de Surco, 15039, Peru |
Gary Nacarino | Department of Maintenance, Infrastructure, Equipment, and General Services Unit, Hospital Nacional Edgardo Rebagliati Martins, EsSalud, Av. Edgardo Rebagliati 490, Jesús María, 15072, Peru |
Jhony A. De La Cruz-Vargas | Instituto de Investigación en Ciencias Biomédicas (INICIB), Ricardo Palma University, Av. Alfredo Benavides 5440, Santiago de Surco, 15039, Peru |
Exoskeletons are crucial for providing intensive and consistent rehabilitation over a longer period and may be able to treat the patient without the presence of the therapist compared to manual therapy. This approach allows for frequent treatment reducing several costs. Therefore, this study aimed to examine the passive elbow rehabilitation of lateral epicondylitis patients, usability, and bioinspired design, to develop a mechatronics system with three rehabilitation positions. Regarding the biomechanical fundamentals of the elbow joint and as an engineering sustain, Computer-Aided Design (CAD) was made, consisting of Finite Element Analysis (FEA), anthropometric, ergonomic, and assembly analysis. The results showed that for the three rehabilitation positions, FEA showed von Mises stress less than the elastic modulus by a 103-factor resulting in no permanent deformation. Position 1, 2, and 3 produced angular displacements of 27°, 16.5° and 31° respectively with a total of 74.5°. An arm exoskeleton for passive rehabilitation of the elbow was developed using a pneumatic cylinder and an AD8832 electromyography (EMG) sensor, capable of detecting the EMG peak point to activate or deactivate the 24 V Arduino relay to flex or extend the elbow based on the positions. A total angular displacement of 74.5° was obtained instead of the simulated version 84.63°, with an error margin of 11.96%. The force during the three rehabilitation positions was 18 N exerted by the air compressor at a 6-bar constant pressure, and due to the use of choke valves.
Arm Exoskeleton; Bio-design; Passive Rehabilitation; Robot-Aided Rehabilitation
The upper limbs are associated with various important roles in daily activities. Therefore, a musculoskeletal and neurological disorder can affect the functions, reducing the patient quality of life, either due to a spinal cord injury, different disorders in motor neurons, or a Cerebrovascular accident (CVA) (Copaci et al., 2019). CVA is one of the main causes of mortality in the world because rehabilitation treatments have a special requirement for one or more therapists due to extended supervision time (Vargas, Cornejo, Correa-López, 2016).
A literature review on
exoskeletons and robotic rehabilitation showed significant advancements in
control and torque estimation. A Model Reference Adaptive Control (MMRAC) has
proven effective for controlling soft robotic exoskeletons, enabling both passive
and assistive control without the need for additional sensors, and showing
robustness against uncertainties (Toro-Ossaba et
al., 2024). Additionally, trend analysis showed that all analyzed
exoskeletons use flexion/extension movements, with aluminum and 3D printing, as
well as PID control algorithms and servomotors (Huamanchahua
et al., 2021). The most common applications are assistance and
rehabilitation (Cornejo et al., 2021).
An adaptive controller based on artificial neural networks demonstrated high
accuracy in gesture detection for wheelchair-mounted exoskeletons (Schabron, Desai, and Yihun, 2021). Twisted String
Actuators (TSAs) have proven effective in lightweight, customizable assistive
devices (Hosseini et al., 2020).
Topological optimization of transtibial prosthetic sockets using Finite Element
Analysis (FEA) demonstrated improvements in stress performance and weight
reduction, although it may also contribute to material fatigue over time (Faadhila, 2022). Compared to manual therapy,
exoskeletons have the potential to provide intensive and consistent
rehabilitation over a longer period (Lo and Xie,
2012).
Other studies focused on the
development and evaluation of innovative medical and rehabilitation devices
aimed at enhancing patient outcomes. For instance, a study designed a
Transforaminal Lumbar Interbody Fusion (TLIF) spine cage using reverse engineering,
followed by simulated compression tests, which demonstrated the capability to
withstand the forces typically encountered in spinal fusion surgeries (Norli et al., 2024). Another study
addressed the challenge of maintaining motivation in pediatric rehabilitation,
particularly for children with Developmental Coordination Disorder (DCD). This
study discussed the iterative co-creation of Matti, a pressure-sensitive
Tangible User Interface (TUI) for exergaming, which has shown potential as a
tool to improve patient engagement and the effectiveness of therapy sessions (Ockerman et al., 2024). These devices can
provide autonomous rehabilitation for patients, potentially reducing healthcare
costs and increasing treatment frequency. The main focus is on the bio-mechatronics
design and manufacturing of an arm exoskeleton with an electro-pneumatic
mechanism tailored for passive rehabilitation.
Passive robotic rehabilitation aims to restore
muscle and joint function by assisting the affected limb without requiring
active participation from the patient (Juarez et
al., 2021). Systems using electromyography (EMG) signals for
rehabilitation are typically categorized as active (Tiboni
et al., 2018), where movements from a functional limb are
mirrored by the affected one (Gull, Bai, and Bak, 2020;
Qassim and Hasan, 2020). However, this study focuses on a passive reflex
rehabilitation method, where both arms are simultaneously engaged, enhancing
recovery through synchronized motion. The classification for the device based
on the type of assistance provided is passive reflex (Maciejasz
et al., 2014) because it aims to assist the motion of the elbow
extremity while mirroring the other arm (Khalid et
al., 2023), The proposed exoskeleton assists in elbow movement, a
critical function in daily tasks, and is designed to improve rehabilitation
outcomes by mirroring the movements of the unaffected limb.
Upper limb
exoskeletons are used to rehabilitate patients with various conditions, such as
strokes (Cervantes et al., 2024),
cerebral palsy (Sandoval et al., 2023),
and neuromuscular diseases, helping to regain mobility of the arms, shoulders,
hands, and elbows (Lo and Xie, 2012). The
elbow is a hinge-type synovial joint located between the humerus with the ulna
and radius formed by the humeral trochlea, the spheroidal condyle, the
trochlear incision, as well as the head of the radius that allow flexion and
extension movements at an angle of up to 170° (Rodriguez
et al., 2022). The proximal radioulnar joint allows pronation and
supination of the forearm (Molina et al.,
2023a). Rotation occurs within a ring formed by the annular ligament and
the radial incision of the ulna, and these movements are essential for everyday
tasks (Rahlin, 2024).
Mechatronic
virtual bio-design and biomechanics are key areas for developing effective arm
exoskeletons for passive elbow rehabilitation (Cornejo,
Cornejo-Aguilar, and Perales-Villarroel, 2019). Virtual bio-design
allows the operation of these devices to be analyzed using computational
models, which has led to significant advances in design (Cornejo, Vargas, and Cornejo-Aguilar, 2020). However,
challenges such as improving comfort, portability, and efficiency remain. Elbow
biomechanics, on the other hand, studies the physical and mechanical principles
of joint movements, providing crucial information for the design of
exoskeletons that adapt to the anatomy and natural movements of the elbow (Molina et al., 2023b). This study
contributes to ongoing efforts by integrating mechatronic virtual bio-design
with elbow biomechanics, creating a low-cost, more effective, and personalized
exoskeleton for passive elbow rehabilitation. The objective is to enhance the
design and performance of exoskeletons through FEA and dynamic simulations,
offering a novel approach to rehabilitation technology.
The
manuscript is organized as follows; Section 2 describes materials and methods
for the bio-mechatronics design of the arm exoskeleton for passive
rehabilitation, including simulation of the control system. Section 3, focuses
on the system manufacturing and integration, while Section 4 describes the test
and results, focusing on the angular displacement and performance of the
system. In Section 5, the manuscript ends with conclusions and further work.
This started by
defining the problem and proposing an innovative solution to the classic
rehabilitation methods, with a robot-aided system intended to be accessible and
easy to use for patients (Cornejo et al.,
2023). To achieve this, the study problem consisted of assessing the
passive elbow rehabilitation of lateral epicondylitis patients and clearly
understanding the symptoms. As a next phase, the project was subjected to
evaluation of the clinical background as well as the usability of bioinspired
design, to develop a mechatronics system with the appropriate materials. The
specific requirements and constraints were then defined to produce a conceptual
design regarding the biomechanical fundamentals of the elbow joint. The
material selection consisted of appropriate elements for the exoskeleton and
the actuators. As an engineering sustain, Computer-Assisted Design (CAD) was
performed, consisting of FEA, anthropometric, and ergonomic, as well as
assembly analysis. Consequently, the mechatronics system design, simulation,
and manufacturing were prepared. This passive elbow system can be used in
rehabilitation and clinical centers, as shown in Figure 1.
2.1. Digital twin
A
digital twin can be described as an integrated multi-physics, multi-scale,
probabilistic simulation of a complex product that uses the best available
physical models and sensor updates to mirror the life of corresponding twin
according to Michael Grieves (Jiang et al.,
2021). Although initially focused on industrial applications, digital
twins have reached the medical sector, for robot-aided rehabilitation, where
the main goal should be relatively simple mechanically. The structure must be
easy to put on and training within the limited space, with an additional
bio-signals tracking system (Falkowski et al.,
2023). Following the digital twin and the design methodology for
rehabilitation robots (Martínez and Z.-Avilés,
2020), the mechanical, electronic, control, and programming system for a
passive elbow rehabilitation exoskeleton station was developed.
Figure 1 Experiment
set-up and apparatus.
2.2. Free
body diagram
Initially, the system was
encharged to transmit force to replicate the flexion and extension of the
elbow, which naturally has a range from 0° to 150° (Martin
and Sanchez, 2013). The functional range of movement is from 30° to 130°
(Felstead and Ricketts, 2017), hence, a
smaller range was considered to not damage the elbow joint, from 30° to 90°. To
reach this point, three rehabilitation positions were considered due to the
limitation of the pneumatic cylinder at 100 mm axial displacement. Each
position produced a 30° angular displacement taking as a reference the position
of the joint. As an initial phase, the free body diagram of the system was
developed to understand how the system will behave as well as the kinematic and
dynamic interactions. Equation (1) describes the geometrical restrictions:
Where represent the lengths of
the link that transmit the force to the revolute joint,
represent the
angle formed with a vertical plane as shown in Figure 2. a., by modelling the
dynamics of the system using the second law of Newton (Rojas-Moreno,
2001). The linear and rotational motion between the first link and the
pneumatic coupling (Figure 2. b.) as well as between the elbow pad and the
second link is described in equations (2), (3), and (4) while the variables are
presented in Table 1:
Table 1 Description of Variables and Model
Parameters
Figure 2 Free body diagram of the system to be developed: a) Geometrical
relationships between the links that transmit force, b) Forces interaction on
the link and the coupling of the pneumatic cylinder, c) Forces interaction
between the link and the revolute joint
2.3. Mechanical design considerations
The
mechanical design was carried out using ISO 7250-3:2015 (ISO, 2015), which standardizes the dimensions of the elbow to
wrist (267 mm), and shoulder to elbow (340 mm). Half of the arm was taken to
make the elbow pads considering ergonomics comfort (Schiele
and Helm, 2006), with one in the arm of 170 mm, and the other in the
forearm of 134 mm. After several simulations, the appropriate link distance was
110 mm to achieve the three rehabilitation positions. For the table support,
the proper distance was 50 mm, and when the user feels uncomfortable,
regulating the chair height can solve this inconvenience. For the actuator,
pneumatic cylinders were considered by offering a high force-to-weight ratio,
being able to be stopped at any time without causing injuries, and requiring
less maintenance, which is desirable for a potentially mobile rehabilitative
exoskeletal system (Karanth P, and Desai, 2022; Burns
et al., 2020; Carvalho, Gopura et
al., 2016; Zhang et al., 2008). As shown in Figure 3, the
hole system can be assembled by the patient putting the table support stable
and then connecting the elbow pad, the link mechanism to the pad, and pneumatic
cylinder. The parts come with a tolerance of 0.15 mm, allowing a correct
coupling, which needs only two ?” screws or the link mechanism. At the same
time, the pneumatic cylinder can be introduced by pressure. To make the system
accessible, polylactic acid (PLA) was considered, as 3D printing technology can
be used to sustain the patient needs in short periods (Demeco
et al., 2023), by customizing the designs and maximizing the
times of fabrication.
Figure 3 System developed
for passive elbow rehabilitation: a) Isometric view, b) Exploded view
2.4. Finite element analysis
In
the analysis, two groups were considered, one for the elbow pads, and the other
for the rehabilitation station. A previous study developed human arm parameters
(Speich, Shao, and Goldfarb, 2005), to
simulate the weight distribution of the forearm and arm using FEA. As a
reference point for the studies, the base of each elbow pad was considered a
fixed geometry. A normal force of 150 N was applied to the face in contact with
the arm in the elbow pads, by taking a median of 10 kg for the human arm mass,
and a scale factor of 1.5 was used for the analysis. Furthermore, the material
selected was ABS from the SolidWorks standard library, with an elastic module
of 2 GPa. The maximum von Mises stress for elbow pad 1 was 56.234 N/m2,
and for elbow pad 2, it was 331.985 N/m2. Under the material elastic
limit, there is no permanent deformation.
The second group of FEM studies was
made to the table support and the linkages. Figure 4a shows that l1
receives a von Mises stress of 135 MPa in the bolted joint, hence, a ?” screw
was considered as the maximum shear stress of 145 MPa according to the
‘Structural Design of Stainless Steel’ (BSSA, 2001).
For the table support, Figure 4b shows a maximum von Mises stress of 8.567 MPa
for the elbow joint, while Figure 4c shows a maximum of 6.287 MPa for the bolted
joint in the pneumatic cylinder coupling. Figure 4d shows a maximum von Miss
stress of 20 MPa in the joint to the elbow pad. However, these values are less
than the elastic module by a 103-factor.
Figure 4 FEA studies for
the elbow pads: a) l1 under a 150 N load with a maximum von Mises
stress of 135 MPa, b) Table support under a 150 N load with a maximum von Mises
stress of 8.567 MPa, c) Pneumatic cylinder coupling under a load of 150 N with
a maximum von Misses stress of 6.287 MPa in the bolted joint, d) Elbow pad
coupling under a load of 150 N with a maximum von Misses stress of 20 MPa.
2.5. Electronic design
The EMG circuit consists of three
electrodes connected to the forearm that capture the electrical activity of the
muscles, connecting to the (1) AD8832 sensor to amplify, filter, and digitize
signals (Zaman et al., 2019). These
signals were sent and processed by the (2) Arduino UNO, to control the (3) 24 V
relay, with a (4) 24 V power supply that powers the relay and valve, while a
separate 5 V supply powers the Arduino, to activate or deactivate the (5)
solenoid valve 5/2, regulating the flow of air. Moreover, the Arduino UNO
connects to an AD8832 configured with a bandwidth of 0.05 Hz to 200 Hz and a
CMRR of 80 dB (Junior et al., 2023).
The ECG electrodes were connected to the IN+ and IN- pins of the AD8832, and
the amplified and filtered signal was sent to the OUT output, connected to an
analog input pin (A0). The microcontroller (ATmega328P) processed and filtered
out high-frequency noise and stabilized the signals. The electrodes detect
voltage changes generated by muscle cells during contraction and transmit to
the EMG circuit. Subsequently, the circuit read with the Arduino program to
determine the readings when opening or closing the hand and moving the arm
unidirectionally (Figure 5).
Figure 5 Circuit diagram
of the electronic system. (1) AD8832, (2) Arduino UNO, (3) 24 V
relay, (4) 24 V power supply, y (5) solenoid valve 5/2
2.6. Pneumatic modeling
A
system model requires knowledge of the physical interactions between forces (Ogata, 2010). MATLAB offers Simscape as a tool to
analyze directly the physical interactions between components (Xavier, Fleming, and Yong, 2020). For this study,
a double-acting pneumatic cylinder was modeled using the fluids library (MATLAB, 2018b), with a reservoir established as a
reference point. The gas properties included a constant of 0.287 kJ/(kg·K),
compressibility factor of 1, reference temperature of 293.15 K, specific heat
at constant pressure of 1 kJ/(kg·K), dynamic viscosity of 18 s·uPa, and a
thermal conductivity of 26 mW/(m·K). The pressure source was in charge of
regulating the flow through the pneumatic circuit, while a 4/3 directional
valve was used to control the flow direction. To connect the load to the
physical system, a translational multibody interface was considered. When the
signal comes from Simulink, it must be converted to a physical signal using the
Simulink-PS, and vice versa. For the EMG input signal simulated as a pulse, it
was converted to a physical signal to activate the directional valve (Figure 6).
2.7. Multibody modeling
Simscape offers multibody links
that act as an interface between CAD software and MATLAB (MATLAB, 2023; MATLAB, 2018a). For this project,
it was used to pass the system developed in SolidWorks to the Multibody
interface. An automatically generated model was made, and the position of the
components was readjusted in Figure 7. The input force for this subsystem was
the pneumatic force due to the EMG signal, which moved the cylindrical joint
between the pneumatic cylinder and the piston rod, limited in the Z prismatic
primitive (Pz) with a lower limit bound of -50 mm, and an upper
limit bound of -25 mm. Gravity was established by the mechanism configuration
block at the Y axis with a vector ([0 -9.80665 0]). Finally, the output of the
signal was the angle between elbow pads 1 and 2 measured using a transform
sensor in radians multiplied by a gain of 180/? in the angular displacement (in
degrees).
Figure 6 The EMG input
activates the pneumatic circuit and changes the position of the 4/3 directional
valve.
2.8. Control system
For
the simulation of the passive elbow exoskeleton rehabilitation system (Figure 8),
MATLAB software was used together with the SimScape library from SIMULINK,
which has been used for modeling prosthetic and rehabilitation systems (McGeehan et al., 2022; Wu et al., 2022; Pitre et al., 2021; Hoh, Chong,
and Etoundi, 2020). By simulating the changes in the position of the 4/3
directional valve, the operation of an electromechanical solenoid button can be
approximated by using a unit step signal that oscillates between -1 and 1. The
arm motion was detected by the EMG sensor and established a peak point. This
changed the position of the valve, extending or contracting the pneumatic
cylinder. Subsequently, the pneumatic force was transferred to the exoskeleton,
and, depending on the rehabilitation position, the angular displacement would
be different.
Figure 8 Control system
for the Passive Rehabilitation Elbow Exoskeleton.
For
all the simulations, the input signal was computed for a period of 20 s with a
duty cycle of 50% and an amplitude of 2. At position 1, the angle between the
elbow pads had a minimum of 58.82° and a maximum of 93.33° with an angular
displacement of 34.51°. At position 2, the angle between the elbow pads had a
minimum of 26.38° and a maximum of 42.3815° with an angular displacement of 16°.
Finally, at position 3, the angle between the elbow pads had a minimum of
0.016° and a maximum of 34.1287° with an angular displacement of 34.11° as
shown in Figure 9. As a total displacement, the partial sum of the positions
was 84.63°, falling within the safety range of motion defined previously. The
simulated EMG signal ranged between 1 and -1 to mimic the elbow joint extension
(1) and flexion (-1). When the patient does flexion, the pneumatic resulting
force activates the pneumatic cylinder to achieve full displacement,
contracting the elbow with a maximum pneumatic force of 18 N. To extend the
joint, the force is gradually reduced over 5 seconds to a minimum of -18 N, as
shown in Figure 10.
Figure 9 Change of angular
position as a function of time.
Figure 10 Pneumatic force is due to the EMG signal as a function of time.
3. System Manufacturing and Integration
3.1. Mechanics Prototyping
As the next phase,
physical integration of the system was made. Initially, the (1) elbow pad, (2)
table support, and (4) linkage mechanism were exported from SolidWorks to
FlashPrint5 for 3D printing (Cornejo et al.,
2022). A previous study showed that PLA has the highest tensile strength
(33.7 MPa) using a layer thickness of 0.27 mm and an infill of 78% (Camargo et al., 2019). To add more
resistance, an infill of 100% was considered as well as tree supports for
corners minor to 20°. By using the base for a correct addition with the heated
bed, favorable results were achieved.
3.2. Electronics Prototyping
A
FESTO 24 V DC supply was used to power the 5/2 solenoid valve. Due to the voltage
and current requirements, a 5 V DC Arduino relay was used to close or open the
circuit when the patient desired. The AD8832 EMG sensor powered with an 18 V
battery, was responsible for receiving the muscular signals of closed and
opened arms, with three electrodes placed in the forearm. The pneumatic
cylinder that exerted motion to the elbow exoskeleton used an air compressor of
220 V connected to the outlet as a power supply. A constant pressure of 6 bar
with choke valves in the cylinder was manually configured to reduce the
velocity of the stem. For the control system, the Arduino UNO prototype board
was used, which received the EMG signal and converted into bits to establish an
activation point at the peak of the muscular activity of the arm, varying
between 500, 800, and 900 for different patients. At this point, the
microcontroller sends 5 V to the relay to close the circuit and the pneumatic
cylinder exerts force to extend or flex the elbow exoskeleton based on the
position of rehabilitation as shown in Figure 11.
3.3. Software Programming
Arduino IDE 2.3.2
software was used due to the advantages compared to previous versions, such as
improvements in stability and performance. Initially, the EMG signal was
defined in the analog pin A0, and the communication port of the relay was
established in pin 8. To determine the readings of the EMG signal, a variable
numReadings was defined to establish the maximum amount of sample.
Subsequently, global definitions for the array that stores the sensor reading,
the index of the reading, the sum of all readings, and the average were
established. The configuration was set up using a communication to the serial
port which had a value of 115200, with EMG_signal being the input, and the
relayPin as an output. Readings were then computed and compared to the EMG_peak.
The relay was activated when the value was greater than the average, or
deactivated when the value was lower, as shown in Figure 12 and Appendix A
Table A.
After the mechanical and electro-pneumatic
prototyping, a test was developed to contrast the angular displacement on the
simulations against the physical system. Initially, the configuration for the
initial test was made by putting the electrodes on the arm with motion. In the
right arm, three electrodes were placed on the forearm, one in the middle
(green), one in two fingers forward than the middle (red), and one below the
forearm (yellow). The left arm, which behaves as the extremity without motion, was
placed on the exoskeleton. Before each test, a calibration of the EMG sensor is
required to activate the system properly. For two patients, the muscular peak
when making a fist received by the AD8832 EMG sensor was 500 and 800 bits
respectively. This suggests that when the activation points are set the same
for both patients, the sensor would not have worked well.
For
position 1 (Figure 13a), the initial angular position had a starting angle of
87° and a final of 60°, with a margin of 6.45% error concerning the initial
design and the digital twin made. For position 2 (Figure 13b), the initial
angle was 25° and the final was 41.5°, indicating a 1.19% error. Finally, for
position 3 (Figure 13c), the initial angle was 7° and the final was 38°, with a
margin of error of 8.57%. This error can be acknowledged due to the cylinder
position when it varies between the start or end point of the pneumatic
actuator. The total angular displacement was 74.5° instead of the simulated
version with 84.63°, and the error margin was 11.96% (Tables 2 and 3). The
pneumatic cylinder exerted a force of 18 N due to the air compressor at a
pressure of 6 kPa. The choke valves were used to reduce the velocity of the
stem and not affect the patient arm.
Figure 13 Comparison between
the three rehabilitation positions: a) Position 1 with an angular displacement
of 27°, b) Position 2 with an angular displacement of 16.5°, c) Position 3 with
an angular displacement of 31°.
Table 2 Test and Simulation Angle achieved for
flexion.
Position |
Tested flexion (°) |
Simulated flexion (°) |
Absolute error (%) |
1 |
87.00 |
93.33 |
6.78 |
2 |
41.50 |
42.38 |
2.08 |
3 |
38.00 |
34.12 |
11.3 |
Table 3 Test and Simulation Angle achieved for
extension.
Position |
Tested extension (°) |
Simulated extension (°) |
Absolute error (%) |
1 |
60.00 |
58.82 |
2.00 |
2 |
25.00 |
26.38 |
5.23 |
3 |
7.00 |
6.00 |
16.67 |
The results obtained can be compared with other related works. Irshaidat developed
an exoskeleton soft robotic arm with various novel pneumatic Muscle Actuators
(pMA) capable of bending (Irshaidat
et al., 2019). To control the bending angle, the
pressure must be controlled by a solenoid valve through a balance of both
tensile forces depending on load and existing air pressure in each muscle of
the actuator arm. For the project developed, the angle was determined by the
rehabilitation position, with constant pressure for the three positions (Rivera
et al., 2020). Burns developed HERCULES, a
three-degree-of-freedom pneumatic upper limb exoskeleton for stroke
rehabilitation (Burns et
al., 2020). The main advantage
of this system is that it manages to rehabilitate the arm by having three
degrees of freedom. However, the dimensions, portability, and costs are higher
than those proposed in this project (Rivera
et al., 2024). Ma et. al. developed a soft wearable
exoskeleton with a pneumatic actuator for assisting the upper limb (Ma et
al., 2020).
The
assistive capability of the exoskeleton is deficient in comparison with
existing devices manufactured with rigid materials. A pneumatic system was used
for the linear transmission of movement instead of electrical rotational motors
that use constant pressure. Therefore, there is no electrical energy
consumption in the compressor until the pressure is below 6 bars. Using
pneumatic linear transmission also keeps a clear working area and can be
stopped under load without any damage (Gopura, Kiguchi, and Bandara,
2011). Compared to rotational transmission, other
exoskeletons that use variable pressures require greater energy consumption. A
proposed exoskeleton design was also assembled by modular pieces (Liu et al., 2021), which allows the reconfiguration of the elbow
pads to adapt in a more personalized way to each patient. This design complies
with safety measures that do not exceed the elastic limit of the material (Zhou et al., 2021). Finally, the open-access technology has great
potential to be used and developed by the same user taking into account
low-cost economic viability (Vargas, Cornejo, and Vargas,
2024).
The authors are
grateful to the Institute of Research in Biomedical Sciences (INICIB),
Universidad Ricardo Palma, Lima, Peru for providing the facilities for 3D
printing as well as guidance for the development of this study.
Conflict
of Interest
The authors declare that there are no
conflicts of interest.
Filename | Description |
---|---|
R1-ME-7197-20240902045003.pdf | --- |
British Stainless Steel Association (BSSA), 2001. Shear and Tension Capacity of Stainless Steel Bolts. Available from: https://bssa.org.uk/bssa_articles/shear-and-tension-capacity-of-stainless-steel-bolts/ April 10, 2024
Burns, M., Zavoda, Z., Nataraj, R., Pochiraju, K., Vinjamuri, R., 2020. HERCULES: A Three Degree-of-Freedom Pneumatic Upper Limb Exoskeleton for Stroke Rehabilitation. In: 2020 42nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), pp. 4959–4962. doi: 10.1109/EMBC44109.2020.9176549
Camargo, J.C., Machado, Á.R., Almeida, E.C., Silva, E.F.M.S., 2019. Mechanical Properties of PLA-Graphene Filament for FDM 3D Printing. The International Journal of Advanced Manufacturing Technology, Volume 103, pp. 2423–2443. doi: 10.1007/s00170-019-03532-5
Carvalho, A.D.D.R., Karanth P, N., Desai, V., 2022. Characterization of Pneumatic Muscle Actuators and Their Implementation on an Elbow Exoskeleton with a Novel Hinge Design. Sensors and Actuators Reports, Volume 4, p. 100109. doi: 10.1016/j.snr.2022.100109
Cervantes, C.M., Sandoval, C., Palomares, R., Arroyo, J.B, Manrique, M.M., Cornejo, J., 2024. Biomedical Anthropometric Evaluation and Conceptual Mechanical Design of Robotic System for Lower Limbs Passive-Rehabilitation on Post-Stroke Patients. Revista de la Facultad de Medicina Humana, Volume 24, pp. 72–81. doi: 10.25176/RFMH.v24i2.6550
Copaci, D., Martin, F., Moreno, L., Blanco, D., 2019. SMA Based Elbow Exoskeleton for Rehabilitation Therapy and Patient Evaluation. IEEE Access, Volume 7, pp. 31473–31484. doi: 10.1109/ACCESS.2019.2902939
Cornejo, J., Barrera, S., Ruiz, C.A.H., Gutierrez, F., Casasnovas, M.O., Kot, L., Solis, M.A., Larenas, R., Castro-Nieny, F., Saavedra, M.R A., Serrezuela, R.R., Londoño, Y.M., Serna, A., Ortega-Aranda, D., Aranda-Miramontes, S., Chang, I., Cardona, M., Carrasquilla-Batista, A., Palomares, R., Rodriguez, R., Parisuaña, R., Bórquez, M., Navarro, O., Sanchez, F., Bonev, I.A., Coulombe, J., Rico, F.M., Treviño-Elizondo, B.L., García-Reyes, H., Sollazzo, A., Dubor, A., Markopoulou, A., Marinis, C.D., Chacin, M., Mora, A., Pérez-Ruiz, M., Ribeiro, A., L'Huillier, E.A., 2023. Industrial, Collaborative and Mobile Robotics in Latin America: Review of Mechatronic Technologies for Advanced Automation. Emerging Science Journal, Volume 7, pp. 1430–1458. doi: 10.28991/ESJ-2023-07-04-025
Cornejo, J., Cornejo-Aguilar, J.A., Vargas, M., Helguero, C.G., Andrade, R.M.D., Torres-Montoya, S., Asensio-Salazar, J., Calle, A.R, Santos, J.M, Damon, A., Quiñones-Hinojosa, A., Quintero-Consuegra, M.D., Umaña, J.P., Gallo-Bernal, S., Briceño, M., Tripodi, P., Sebastian, R., Perales-Villarroel, P., Cruz-Ku, G.F.L., McKenzie, T., Arruarana, V.S., Ji, J., Zuluaga, L., Haehn, D.A., Paoli, A., Villa, J.C., Martinez, R., Gonzalez, C., Grossmann, R.J., Escalona, G., Cinelli, I., Russomano, T., 2022. Anatomical Engineering and 3D Printing for Surgery and Medical Devices: International Review and Future Exponential Innovations. BioMed Research International, Volume 2022(1), p. 6797745. doi: 10.1155/2022/6797745
Cornejo, J., Cornejo Aguilar, J.A., Perales Villarroel, J.P., 2019. International Innovations in Medical Robotics to Improve Patient Management in Peru. Journal of the Faculty of Human Medicine, Volume 19(4), pp. 105–113. doi: 10.25176/RFMH.v19i4.2349
Cornejo, J., Huamanchahua, D., Huamán-Vizconde, S., Terrazas-Rodas, D., Sierra-Huertas, J., Janampa-Espinoza, A., Gonzáles, J., Cardona, M., 2021. Mechatronic Exoskeleton Systems for Supporting the Biomechanics of Shoulder-Elbow-Wrist: An Innovative Review. In: 2021 IEEE International IOT, Electronics and Mechatronics Conference (IEMTRONICS), pp. 1–9. doi: 10.1109/IEMTRONICS52119.2021.9422660
Cornejo, J., Vargas, M. & Cornejo-Aguilar, J.A., 2020. Robotics and Biomedical Innovative Applications in Public Health during the COVID-19 Pandemic: Innovative Applications of Robotics and Biomedicine in Public Health During the COVID-19 Pandemic. doi: 10.25176/RFMH.v20i4.3042
Demeco, A., Foresti, R., Frizziero, A., Daracchi, N., Renzi, F., Rovellini, M., Salerno, A., Martini, C., Pelizzari, L., Costantino, C., 2023. The Upper Limb Orthosis in the Rehabilitation of Stroke Patients: The Role of 3D Printing. Bioengineering, Volume 10(11), p. 1256. doi: 10.3390/bioengineering10111256
Faadhila, A., Rahman, S.F., Whulanza, Y., Supriadi, S., Tampubolon, J.Y., Wicaksana, S.I., Rahyussalim, A.J., Kurniawati, T., Abdullah, A.H., 2022. Design of a Transforaminal Lumbar Interbody Fusion (TLIF) Spine Cage. International Journal of Technology, Volume 13(8), pp. 291–319. doi: 10.14716/ijtech.v13i8.6152
Falkowski, P., Osiak, T., Wilk, J., Prokopiuk, N., Leczkowski, B., Pilat, Z., Rzymkowski, C., 2023. Study on the Applicability of Digital Twins for Home Remote Motor Rehabilitation. Sensors, Volume 23. p. 911, doi: 10.3390/s23020911
Felstead, A.J., Ricketts, D., 2017. Biomechanics of the Shoulder and Elbow. Orthopaedics and Trauma, Volume 31(5), pp. 300–305. doi: 10.1016/j.mporth.2017.07.004
Gopura, R.A.R.C., Bandara, D.S.V., Kiguchi, K., Mann, G.K I., 2016. Developments in Hardware Systems of Active Upper-Limb Exoskeleton Robots: A review. Robotics and Autonomous Systems, Volume 75, pp. 203–220. doi: 10.1016/j.robot.2015.10.001
Gopura, R.A.R.C., Kiguchi, K., Bandara, D.S.V., 2011. A Brief Review on Upper Extremity Robotic Exoskeleton Systems. In: 2011 6th International Conference on Industrial and Information Systems, pp. 346–351. doi: 10.1109/ICIINFS.2011.6038092
Gull, M.A., Bai, S., Bak, T., 2020. A Review on Design of Upper Limb Exoskeletons. Robotics, Volume 9(1), p. 16. doi: 10.3390/robotics9010016
Hoh, S., Chong, J. Etoundi, A.C., 2020. Design of a Virtual Testing Platform for assessing Prosthetic Knee Joints. In: 2020 5th International Conference on Advanced Robotics and Mechatronics (ICARM), pp. 576–581. doi: 10.1109/ICARM49381.2020.9195275
Hosseini, M., Meattini, R., San-Millan, A., Palli, G., Melchiorri, C., Paik, J., 2020. A sEMG-Driven Soft ExoSuit Based on Twisted String Actuators for Elbow Assistive Applications. IEEE Robotics and Automation Letters, Volume 5(3), pp. 4094–4101. doi: 10.1109/LRA.2020.2988152
Huamanchahua, D., Castañeda-Vásquez, C., Vásquez-Espinoza, A. Muñoz-Zevallos, A, 2021. Robotic Devices Types Exoskeletons for Elbow Rehabilitation: A Technological Review. In: 2021 IEEE 12th Annual Ubiquitous Computing, Electronics & Mobile Communication Conference (UEMCON), pp. 0791–0796. doi: 10.1109/UEMCON53757.2021.9666652
Irshaidat, M.A.E., Soufian, M., Al-Ibadi, A., Nefti-Meziani, S., 2019. A Novel Elbow Pneumatic Muscle Actuator for Exoskeleton Arm in Post-Stroke Rehabilitation. In: 2019 2nd IEEE International Conference on Soft Robotics (RoboSoft), pp. 630–635. doi: 10.1109/ROBOSOFT.2019.8722813
Jiang, Y., Yin, S., Li, K., Luo, H., Kaynak, O., 2021. Industrial Applications of Digital Twins. Philosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences, Volume 379(2207), p. 20200360. doi: 10.1098/rsta.2020.0360
Juarez, J.A., Palomares, R., Cornejo, J. Bustinza, R.R., 2021. Biomedical Mechatronic Device iTakuna: Smart Embedded System for Electric Wheelchair used by Paraplegic Patients. In: 2021 IEEE XXVIII International Conference on Electronics, Electrical Engineering and Computing (INTERCON), pp. 1–4. doi: 10.1109/INTERCON52678.2021.9532609
Junior, J.J.A.M., Campos, D.P., Biassio, L.C.D.A.V.D., Passos, P.C., Júnior, P.B., Lazzaretti, A.E. Krueger, E., 2023. AD8232 to Biopotentials Sensors: Open Source Project and Benchmark. Electronics, Volume 12(4), p. 833. doi: 10.3390/electronics12040833
Khalid, S., Alnajjar, F., Gochoo, M., Renawi, A., Shimoda, S., 2023. Robotic Assistive and Rehabilitation Devices Leading to Motor Recovery in Upper Limb: A Systematic Review. Disability and Rehabilitation: Assistive Technology, Volume 18(5), pp. 658–672. doi: 10.1080/17483107.2021.1906960
Liu, Q., Liu, Y., Zhu, C., Guo, X., Meng, W., Ai, Q., Hu, J., 2021. Design and Control of a Reconfigurable Upper Limb Rehabilitation Exoskeleton with Soft Modular Joints. IEEE Access, Volume 9, pp. 166815–166824. doi: 10.1109/ACCESS.2021.3136242
Lo, H.S., Xie, S.Q., 2012. Exoskeleton Robots for Upper-Limb Rehabilitation: State of the Art and Future Prospects. Medical Engineering & Physics, Volume 34, pp. 261–268. doi: 10.1016/j.medengphy.2011.10.004
Ma, J., Chen, D., Liu, Z., Wang, M., 2020. A Soft Wearable Exoskeleton with Pneumatic Actuator for Assisting Upper Limb. In: 2020 IEEE International Conference on Real-time Computing and Robotics (RCAR), pp. 99–104. doi: 10.1109/RCAR49640.2020.9303269
Maciejasz, P., Eschweiler, J., Gerlach-Hahn, K., Jansen-Troy, A., Leonhardt, S., 2014. A Survey on Robotic Devices for Upper Limb Rehabilitation. Journal of NeuroEngineering and Rehabilitation, Volume 11, pp. 1–29. doi: 10.1186/1743-0003-11-3
Martin, S., Sanchez, E., 2013. Anatomy and Biomechanics of the Elbow Joint. In: Seminars in Musculoskeletal Radiology, Volume 17(5), pp. 429–436. doi: 10.1055/s-0033-1361587
Martínez, G.M.C., Avilés, L.A.Z., 2020. Design Methodology for Rehabilitation Robots: Application in an Exoskeleton for Upper Limb Rehabilitation. Applied Sciences, Volume 10(16), p. 5459. doi: 10.3390/app10165459
MATLAB, 2018a. Enable Simscape Multibody Link Plugin in SolidWorks - MATLAB & Simulink, https://www.mathworks.com/help/smlink/ref/linking-and-unlinking-simmechanics-link-software-with-solidworks.html
MATLAB, 2018b. Pneumatic Actuation Circuit - MATLAB & Simulink, https://www.mathworks.com/help/simscape/ug/pneumatic-actuation-circuit.html
MATLAB, 2023. Import SolidWorks Assemblies into Simscape Multibody, https://www.mathworks.com/videos/import-solidworks-assemblies-into-simscape-multibody-1701670328083.html
McGeehan, M.A., Adamczyk, P.G., Nichols, K.M., Hahn, M.E., 2022. A Simulation-Based Analysis of the Effects of Variable Prosthesis Stiffness on Interface Dynamics Between the Prosthetic Socket and Residual Limb. Journal of Rehabilitation and Assistive Technologies Engineering, Volume 9, p. 20556683221111986. doi: 10.1177/20556683221111986
Molina, D.A.Z, Cabrejos, R.J.S., Cornejo, J., Manrique, M.M., Rodríguez, R., Palomares, R., 2023a. Mechatronics Design and Bio-Motion Simulation of Trans-radial Arm Prosthesis Controlled by EMG Signals. In: 2023 IEEE Colombian Caribbean Conference (C3), pp. 1–7. doi: 10.1109/C358072.2023.10436316
Molina, D.A.Z, Cabrejos, R.J.S, Manrique, M.F.M., Bustinza, R.R.R., Aguilar, J.L.C, Orihuela, R.J.P, 2023b. Modelado Geométrico y Simulación Cinemática de Prótesis de Mano para Pacientes Adultos con Amputación Transradial (Geometric Modeling and Kinematic Simulation of Hand Prostheses for Adult Patients with Transradial Amputation). Perfiles de Ingeniería, Volume 19(20), pp. 179–200. doi: 10.31381/perfilesingenieria.v19i20.6020
Norli, M.H.M., Sukimi, A.K.A., Ramlee, M.H., Mahmud, J. Abdullah, A.H., 2024. Static Structural Analysis on Different Topology Optimization Transtibial Prosthetic Socket Leg. International Journal of Technology, Volume 15(2), pp. 455–462. doi: 10.14716/ijtech.v15i2.6711
Ockerman, J., Octavia, J.R., Joundi, J., Penders, A., Bar-On, L., Saldien, J., 2024. Matti: Tangible User Interface for Engaging Patients in Physical Therapy Towards a Motivating Rehabilitation. International Journal of Technology, Volume 15, pp. 697–708. doi: 10.14716/ijtech.v15i3.6118
Ogata, K., 2010. Modern Control Engineering. 5th edition. Pearson Education
Pitre, S., Curtin, B., Pena, P., Rouse, C., Joseph, E., Hooper, J., Tekes, A., 2022. Preliminary Design and Experimental Studies of a Compliant Knee Joint for Pediatric Above Knee Amputees. In: ASME International Mechanical Engineering Congress and Exposition, Volume 85598, p. V005T05A072, American Society of Mechanical Engineers. doi: 10.1115/IMECE2021-73655
Qassim, H.M., Hasan, W.Z.W., 2020. A Review on Upper Limb Rehabilitation Robots. Applied Sciences, Volume 10(19), p. 6976. doi: 10.3390/app10196976
Rahlin, M., 2024. Physical Therapy for Children with Cerebral Palsy: An Evidence-Based Approach, Routledge, Taylor & Francis Group, New York, USA,
Rivera, M.V., Cornejo, J., Huallpayunca, K., Diaz, A.B., Ortiz-Benique, Z.N., Reina, A.D., Jamanca Lino, G., Ticllacuri, V., 2020. Human Space Medicine: Physiological Performance and Countermeasures to Improve Astronaut Health. Journal of the Faculty of Human Medicine, Volume 20(2), pp. 303–314. doi: 10.25176/RFMH.v20i2.2920
Rivera, M.V., Vargas, M., Cornejo, J., Plascencia, P.V., Guillen, K., Maquera, E., Cornejo, J., Russomano, T., Cinelli, I., 2024. Space Nursing for the Future Management of Astronaut Health in other Planets: A Literature Review. The Open Nursing Journal, Volume 18(1), p. e18744346289848. doi: 10.2174/0118744346289848240328074640
Rodriguez, V., Sanchez, L., Palomares, R. & Cornejo, J. Ergonomic Bio-Design and Motion Simulation of a Mechatronic Orthosis System for Elbow Rehabilitation. In: 2022 IEEE XXIX International Conference on Electronics, Electrical Engineering and Computing (INTERCON), pp. 1–4. doi: 10.1109/INTERCON55795.2022.9870083
Rojas-Moreno, A., 2001. Advanced Control: Design and Real-Time Applications. National University of Engineering. Independent Publication
Sandoval, C., Martel, C., Palomares, R., Arroyo, J.B., Manrique, M.F.M. Cornejo, J., 2023 Conceptual Mechatronic Design of Ankle-foot Exoskeleton System for Assisted Rehabilitation of Pediatric Patients with Spastic Cerebral Palsy. In: 2023 IEEE MIT Undergraduate Research Technology Conference (URTC), pp. 1–5. doi: 10.1109/URTC60662.2023.10534922
Schabron, B., Desai, J. & Yihun, Y., 2021. Wheelchair-Mounted Upper Limb Robotic Exoskeleton with Adaptive Controller for Activities of Daily Living. Sensors, Volume 21(17), p. 5738. doi: 10.3390/s21175738
Schiele, A. & Helm, F.C.V.D., 2006. Kinematic Design to Improve Ergonomics in Human Machine Interaction. IEEE Transactions on Neural Systems and Rehabilitation Engineering, Volume 14, pp. 456–469. doi: 10.1109/TNSRE.2006.881565
Speich, J.E., Shao, L., Goldfarb, M., 2005.Modeling the Human Hand as It Interacts with a Telemanipulation System. Mechatronics, Volume 15, pp. 1127-1142. doi: 10.1016/j.mechatronics.2005.06.001
The International Organization for Standardization (ISO), 2015. ISO 7250-3:2015. Basic Human Body Measurements for Technological Design Part 3: Worldwide and Regional Design Ranges for Use in Product Standards. https://www.iso.org/es/contents/data/standard/06/42/64237.html?browse=ics
Tiboni, M., Borboni, A., Faglia, R., Pellegrini, N., 2018. Robotics Rehabilitation of the Elbow Based on Surface Electromyography Signals. Advances in Mechanical Engineering, Volume 10(2), p. 1687814018754590. doi: 10.1177/1687814018754590
Toro-Ossaba, A., Tejada, J.C., Rúa, S., Núñez, J.D., Peña, A., 2024. Myoelectric Model Reference Adaptive Control with Adaptive Kalman Filter for a Soft Elbow Exoskeleton. Control Engineering Practice, Volume 142, p. 105774. doi: 10.1016/j.conengprac.2023.105774
Vargas, M., Cornejo, J. Correa-López, L.E., 2016. Ingeniería Biomédica: La Revolución Tecnológica para el Futuro del Sistema de Salud Peruano. Journal of the Faculty of Human Medicine, Volume 16(2). doi: 10.25176/RFMH.v16.n3.659
Vargas, M., Cornejo, J., Vargas, J.A.D.L.C., 2024. Health without Borders: INICIB-URP Leading Medical Education through Bioengineering and Mechatronic Technologies: Salud sin Fronteras: INICIB-URP liderando la Educación Médica a través de Bioingeniería y Tecnologías Mecatrónicas. Journal of the Faculty of Human Medicine, Volume 24(3), pp. 7–11. doi: 10.25176/RFMH.v24i3.6701
Wu, Z., Chen, Y., Geng, Y., Wang, X., Xuan, B., 2022. Model-Free Robust Adaptive Integral Sliding Mode Impedance Control of Knee–Ankle–Toe Active Transfemoral Prosthesis. The International Journal of Medical Robotics and Computer-Assisted Surgery, Volume 18(3), p. e2378. doi: 10.1002/rcs.2378
Xavier, M.S., Fleming, A.J., Yong, Y.K. Modelling and Simulation of Pneumatic Sources for Soft Robotic Applications. In: 2020 IEEE/ASME International Conference on Advanced Intelligent Mechatronics (AIM), pp. 916–921. doi: 10.1109/AIM43001.2020.9158802
Zaman, M.I.U., Tabassum, S., Ullah. M.S., Rahaman, A., Nahar, S., Islam, A.K.M.M., 2019. Towards IoT and ML Driven Cardiac Status Prediction System. In: 2019 1st International Conference on Advances in Science, Engineering and Robotics Technology (ICASERT), IEEE, pp. 1–6. doi: 10.1109/ICASERT.2019.8934639
Zhang, J.F., Yang, C.J., Chen, Y., Zhang, Y., Dong, Y.M., 2008. Modeling and Control of a Curved Pneumatic Muscle Actuator for Wearable Elbow Exoskeleton. Mechatronics, Volume 18(8), pp. 448–457. doi: 10.1016/j.mechatronics.2008.02.006
Zhou, X., Liu, G., Han, B., Wu, L., Li, H., 2021. Design of a Human Lower Limbs Exoskeleton for Biomechanical Energy Harvesting and Assist Walking. Energy Technology, Volume 9(1), p. 2000726. doi: 10.1002/ente.202000726